Operating Mode
: |
Corona Aerosol Detector |
|
Mobile Phase Flow
Rate : |
Up to 2 mL/min |
|
Wettable Surfaces : |
316
stainless steel, Teflon, glass, chromium [also o-ring materials] |
|
Full Scale Ouput
Range : |
1 pA to 500
pA in 1-2-5 sequence |
|
Filter Time
Constants : |
None, Low,
Medium, High |
|
Noise Specification : |
< 750
fA peak to peak (20% methanol/80% water) |
|
Signal Output : |
0 V to 1 V
DC |
|
Output Resolution : |
0.12 µV at
1 V full scale (24 bit bipolar) |
|
Maximum Sampling Rate : |
3.75
samples/sec |
|
Output Offset : |
±
50% of the
selected current range in 1% steps |
|
Event
Marker : |
Triggered by keypad,
timed operation or RS232 control, width, height, polarity and channel
selectable |
|
Autozero : |
Triggered
by front panel keypad, rear panel contact closure, timed operation or
RS232 control |
|
Function Keys : |
Autozero,
Event Marker, Gas On/Off |
|
RS232 Interface Capability : |
Full
parametric instrument control |
|
Method Storage : |
Up to 25
methods (any combination of Fixed Mode and Timeline) |
|
Settable Temperature Range : |
Ambient +5°
to 60°C (Thermal Organizer Option) |
|
Temperature Stability : |
<
±
0.5°C
(Thermal Organizer Option) |
|
Warm-up Time : |
< 30
minutes (typically) (Thermal Organizer Option) |
|
Temperature Accuracy : |
<
±
0.5°C (Thermal Organizer Option) |
| Specifications subject to