Number
of Channels
: |
4 |
|
Electrode Configuration : |
2 or 3 |
|
Operational Modes : |
Amperometric, Voltammetric, Temperature |
|
Input Signal Polarity : |
Bipolar |
|
Input Signal Configuration : |
Single ended |
|
Range Settings |
|
|
For Amperometric Mode |
± 1 nA
to ±1 mA |
|
For Voltammetric Mode |
± 5000
mV |
|
For Temperature Mode |
± 100°C |
|
Measurement Accuracy: |
|
|
For Amperometric/Voltammetric Mode |
0.1% x
full scale (10 fA in nA range) |
|
For Temperature Mode |
± 1% |
|
Activation Voltage Range |
-2000
mV to +2000 mV |
|
Input Impedance |
> 1012
Ohm |
|
Baseline Noise |
10-6
x full scale of selected range |
|
Resolution |
10-5
x full scale of selected range |
|
Operating System – host computer |
Windows® 2000 or later |
|
Power Supply |
9 to 30
VDC 10 W unregulated |
|
Instrument Dimensions |
35 mm x 166 mm x 125
mm |
|
Weight |
0.65 kg |
| Specifications subject to